Validation of the ReeVue® and the CardioCoach®CO2 for Measuring Resting Energy Expenditure

PURPOSE: To examine the validity of the ReeVue™ (RV) and CardioCoachCO2™ (CC) metabolic systems to measure oxygen consumption by comparison to a reference criterion, the MedicalGraphics CardiO2/CP™ (MG).

RESULTS: There were no statistical differences between the two MG tests. Therefore, MG VO2 values were averaged and compared to the RV and CC. No significant differences were found between the RV (202±45 mLO2/min), CC (209±51 mLO2/min), and MG (226±57 mLO2/min) systems.

CONCLUSIONS: These findings indicate that the ReeVue™ and CardioCoachCO2™ metabolic systems are comparable devices to the selected reference criterion for measuring oxygen consumption to assess resting energy expenditure.

Sun S, et al.  Medicine & Science in Sports & Exercise: May 2009 – Volume 41 – Issue 5 – p 42

Was this article helpful?

Have Additional Questions?

Our KORR Medical Technology specialists are trained to answer any questions you may have. Call us at 1-877-859-3007, or click one of the options below.

Help Center Email Us Call

Replacement Supplies

Browse Supplies
Save & Share Cart
Your Shopping Cart will be saved and you'll be given a link. You, or anyone with the link, can use it to retrieve your Cart at any time.
Back Save & Share Cart
Your Shopping Cart will be saved with Product pictures and information, and Cart Totals. Then send it to yourself, or a friend, with a link to retrieve it at any time.
Your cart email sent successfully :)

The European Union’s ePrivacy Directive (often referred to as the ‘cookie law’) and General Data Protection Regulation (GDPR) places requirements to provide information about, and gain consent for the use of cookies. This site uses cookies. By continuing to use this website, you agree to their use. To find out more, including how to control cookies, see here